FDA Recall Terminated

MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1

Recall: Z-0344-2023 · Initiated November 18, 2022

Recall

Recall Number
Z-0344-2023
Event Number
91178
Firm
Mobius Imaging, LLC
FEI Number
3010151377
Product Code
JAK
Status
Terminated
Root Cause
Other
Initiated
November 18, 2022
Posted
December 6, 2022
Terminated
February 26, 2025
Address
2 Shaker Rd, Ste F100, Shirley, MA, 01464-2535

Description

MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1

Reason

Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.

Action

On November 18, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. Customers were informed that a service representative will contact their facility to schedule a time to replace the batteries in the affected devices. The customer does not need to take additional action while waiting for the battery preplacement. If you have any questions or concerns regarding this update, please contact [email protected] or call Stryker Technical Support at 978 386 9619.

Distribution

Domestic distribution to consignees in Colorado, Tennessee, and Utah.

Quantity

3