FDA Recall Terminated

SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Version 4.0.1.14 Manufactured & Distributed 12/29/2006. Version 4.0.1.15 Manufactured & Distributed 05/02/2007. Version 4.0.1.16 Manufactured & Distributed 08/22/2008. Used in a medical research or clinical laboratory setting by knowledgeable, trained and experienced personnel.

Recall: Z-0344-2011 · Initiated February 16, 2010

Recall

Recall Number
Z-0344-2011
Event Number
56863
Firm
SCC Soft Computer
FEI Number
1000307651
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
February 16, 2010
Posted
November 10, 2010
Terminated
July 17, 2012
Address
5400 Tech Data Drive, Clearwater, FL, 33760

Description

SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Version 4.0.1.14 Manufactured & Distributed 12/29/2006. Version 4.0.1.15 Manufactured & Distributed 05/02/2007. Version 4.0.1.16 Manufactured & Distributed 08/22/2008. Used in a medical research or clinical laboratory setting by knowledgeable, trained and experienced personnel.

Reason

When a test is being verified, the wrong (expired) ranges may be associated to that test result, causing an incorrect flag to be posted for that test result.

Action

Clients were notified via the proprietary communication software on 02/16/2010. The Notification Verbiage identified the affected product and informed clients of the potential hazard, root cause, corrective action, and requested action. Clients were instructed to acknowledge receipt of the notification. If the client wishes to receive the correction, they should document a request to load the software.

Distribution

Nationwide Distribution - US and Puerto Rico.

Quantity

31