FDA Recall Terminated

Copper NiTi Preformed Archwires; Copper NiTi 40 BAF .016X.022 LO Pk10, Part No. 211-0952. Intended to provide tooth movement.

Recall: Z-0343-2016 · Initiated November 4, 2015

Recall

Recall Number
Z-0343-2016
Event Number
72590
Firm
Ormco/Sybronendo
FEI Number
2016150
Product Code
DZC
Status
Terminated
Root Cause
Employee error
Initiated
November 4, 2015
Posted
November 30, 2015
Terminated
January 19, 2016
Address
1332 S Lone Hill Ave, Glendora, CA, 91740-5339

Description

Copper NiTi Preformed Archwires; Copper NiTi 40 BAF .016X.022 LO Pk10, Part No. 211-0952. Intended to provide tooth movement.

Reason

Ormco Corporation is recalling Copper NiTi Preformed Archwires because the product transformation temperature as listed on the labeling is incorrect.

Action

A recall letter dated 11/5/15 was sent to all their customers who purchased the Copper NiTi Preformed Archwires because it has been brought to Ormco's attention that the product transformation temperature as listed on the labeling is incorrect. The wire is labeled as having 40 degrees Celsius transformation temperature, but the actual wire has a transformation temperature of 29 degrees Celsius. The letter provides the customers with the problems identified and the actions to be taken. Customers are instructed to complete and fax back the enclosed acknowledgement form.

Distribution

Distributed US (nationwide) and the countries of Canada, Japan, Mexico, Russia, and Peru.

Quantity

161 units