CAIRE Liberator 30, MODEL 13337403, Liquid Oxygen System Unit
Recall
- Recall Number
- Z-0338-2023
- Event Number
- 90999
- Firm
- Caire, Inc.
- FEI Number
- 3004972304
- Product Code
- BYJ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- October 10, 2022
- Posted
- December 2, 2022
- Terminated
- August 19, 2024
- Address
- 2200 Airport Industrial Dr, Ste 500, Ball Ground, GA, 30107-4686
Description
CAIRE Liberator 30, MODEL 13337403, Liquid Oxygen System Unit
An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
The firm issued an URGENT: MEDICAL DEVICE RECALL notice its consignees be email and telephone. The notice explained the issue and requested the removal of the affected products from the field and return to CAIRE. CAIRE will immediately replace the above serial numbers with new Liberator devices. Please contact CAIRE Customer Service via phone at 1-800-482-2473 or 770-721-7700 or via email at [email protected] to arrange for the return and replacement of the above serial numbers.
US Nationwide - Worldwide Distribution: CA, NC, OH, AZ, and Canada, Chile, Colombia, and Germany
5 devices