FDA Recall Terminated

Boston Scientific iLab Ultrasound Imaging System, models 120INS and 240INS. Its intended use is for ultrasound examination of intravascular pathology. Manufactured by Boston Scientific Corporation, Fremont, CA.

Recall: Z-0338-2009 · Initiated September 5, 2008

Recall

Recall Number
Z-0338-2009
Event Number
49420
Firm
Boston Scientific Corporation
FEI Number
2939204
Product Code
IYO
Status
Terminated
Root Cause
Process control
Initiated
September 5, 2008
Posted
December 16, 2008
Terminated
May 18, 2009
Address
47900 Bayside Pkwy, Fremont, CA, 94538-6515

Description

Boston Scientific iLab Ultrasound Imaging System, models 120INS and 240INS. Its intended use is for ultrasound examination of intravascular pathology. Manufactured by Boston Scientific Corporation, Fremont, CA.

Reason

Computer malware: Products may be infected with a worm (computer malware) which could infect a computer network to which it may be connected.

Action

Customer letters sent on Friday, Sept 5 via FedEx, 2008 to all customers. This letter will be followed up by a customer service visit. Package delivery will be tracked, as well as Reply Verification tracking forms.For further information, please contact firm at 508-652-5594.

Distribution

World-wide distribution including USA and countries of Europe, Middle East, Africa, and Japan.

Quantity

357 units