FDA Recall
Terminated
Hill-Rom brand Century+ bed; model P1400.
Recall: Z-0338-04
·
Initiated December 26, 2003
Recall
- Recall Number
- Z-0338-04
- Event Number
- 28009
- Firm
- Hill-Rom, Inc.
- FEI Number
- 1824206
- Product Code
- FNL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 26, 2003
- Posted
- January 10, 2004
- Terminated
- May 13, 2004
- Address
- 125 E Pearl St, Batesville, IN, 47006
Description
Hill-Rom brand Century+ bed; model P1400.
Reason
If fluid is spilled on the siderail or the bed control areas, the bed may change positions without user input.
Action
A recall letter dated December 26, 2003 was sent to each customer. Customers are asked to follow instructions in the user manual, if there is a fluid spill onto a unit, until Hill-Rom can inspect each unit to determine whether it is affected by the problem.
Distribution
United States, Canada , Panama and Taiwan.
Quantity
222