FDA Recall Terminated

Terumo Advanced Perfusion System 1 Control Unit; Catalog number 801046.

Recall: Z-0335-2007 · Initiated October 12, 2004

Recall

Recall Number
Z-0335-2007
Event Number
36396
Firm
Terumo Cardiovascular Systems Corp
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Other
Initiated
October 12, 2004
Posted
December 30, 2006
Terminated
February 1, 2007
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586

Description

Terumo Advanced Perfusion System 1 Control Unit; Catalog number 801046.

Reason

Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.

Action

U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.

Distribution

Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.

Quantity

4,740 for all affected products