FDA Recall
Terminated
3D TOP Ceiling Stand. X-Ray System model number 8773673
Recall: Z-0335-06
·
Initiated November 30, 2005
Recall
- Recall Number
- Z-0335-06
- Event Number
- 34191
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- IYB
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 30, 2005
- Posted
- December 31, 2005
- Terminated
- July 19, 2006
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355-1406
Description
3D TOP Ceiling Stand. X-Ray System model number 8773673
Reason
screws on the rollers may become loose
Action
The recalling firm has issued a field correction per Update Instructions AX004/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees and checkthe screws on the rollers. It also provides roller bearing sleeves as a secondary safety mechanism to prevent the rollers from detaching.
Distribution
The products were shipped to hospitals and medical facilities nationwide.
Quantity
460 units