FDA Recall Terminated

3D TOP Ceiling Stand. X-Ray System model number 8773673

Recall: Z-0335-06 · Initiated November 30, 2005

Recall

Recall Number
Z-0335-06
Event Number
34191
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IYB
Status
Terminated
Root Cause
Other
Initiated
November 30, 2005
Posted
December 31, 2005
Terminated
July 19, 2006
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406

Description

3D TOP Ceiling Stand. X-Ray System model number 8773673

Reason

screws on the rollers may become loose

Action

The recalling firm has issued a field correction per Update Instructions AX004/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees and checkthe screws on the rollers. It also provides roller bearing sleeves as a secondary safety mechanism to prevent the rollers from detaching.

Distribution

The products were shipped to hospitals and medical facilities nationwide.

Quantity

460 units