FDA Recall Terminated

Superstat Modified Collagen Hemostatic Sponge, 50mm Disc, 3.0%, Part# 9200-30. For use when hemostasis is desired along suture lines, in diffusely bleeding sites, in incisions, dissections, and around arterial or venous anastomoses. Indicated in oozing bleeding, either capillary or venous, in almost all types of incisions or wounds, and as an adjunct to conventional closure techniques

Recall: Z-0334-2013 · Initiated October 24, 2012

Recall

Recall Number
Z-0334-2013
Event Number
63581
Firm
Superstat Corp
FEI Number
3001452136
Product Code
LMF
Status
Terminated
Root Cause
Process control
Initiated
October 24, 2012
Posted
November 16, 2012
Terminated
February 6, 2013
Address
2015 E University Dr, Rancho Dominguez, CA, 90220-6411

Description

Superstat Modified Collagen Hemostatic Sponge, 50mm Disc, 3.0%, Part# 9200-30. For use when hemostasis is desired along suture lines, in diffusely bleeding sites, in incisions, dissections, and around arterial or venous anastomoses. Indicated in oozing bleeding, either capillary or venous, in almost all types of incisions or wounds, and as an adjunct to conventional closure techniques

Reason

The recall was initiated because Superstat Corporation has confirmed that no documentation are available to verify that quarterly dose sterility audits were performed for the Superstat Modified Collagen Hemostatic Sponge.

Action

Superstat initially contacted customers by telephone and sent an Urgent - Device Recall Notification Letter on October 24, 2012, to all affected customers. The recall letter provided the reason for recall was due to inadequate documentation of quarterly dose sterility audits. The recall includes all lots and part numbers of Superstat produced between 2002 and 2012. Customers are instructed to complete and return the enclosed response form as soon as possible. Customers with questions are instructed to call customer service at (800) 487-3786. For questions regarding this recall call 310-605-1655.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, Thailand, Mexico, Canada, Pakistan, Israel, Greece, Serbia, Canda, Isle of Man, Nova Scotia, and Singapore.

Quantity

26,422 units