FDA Recall Open, Classified

Boston Scientific AUTOLITH TOUCH Bipolar Electrohydraulic Lithotripter, REF M005466800. The device is designed to be used with Northgate Technologies Inc. bipolar disposable EHL probes for the fragmentation of calculi.

Recall: Z-0333-2022 · Initiated October 25, 2021

Recall

Recall Number
Z-0333-2022
Event Number
88944
Firm
Northgate Technologies, Inc.
FEI Number
1450997
Product Code
FFK
Status
Open, Classified
Root Cause
Software design (manufacturing process)
Initiated
October 25, 2021
Address
1591 Scottsdale Ct, Elgin, IL, 60123-9361

Description

Boston Scientific AUTOLITH TOUCH Bipolar Electrohydraulic Lithotripter, REF M005466800. The device is designed to be used with Northgate Technologies Inc. bipolar disposable EHL probes for the fragmentation of calculi.

Reason

The languages for Russian, Romanian, Slovak, and Czech have the power settings for high and low switched in the device.

Action

The recalling firm, Northgate Technologies Inc., issued a recall letter dated 10/25/2021 to their customer explaining the reason for recall. The customer was requested to determine the location of the affected devices. If the device is/could be used in one of the affected languages, the device is to be returned so the software can be updated to address the text mix-up. If the device is not or will not be operated in the affected languages, the device does not need to be returned immediately and the software will be upgraded during the next servicing to address the mix-up. Complete the enclosed Recall Acknowledgment Form indicating the serial numbers, distribution dates, and whether the device needs correction now or later and email it to: [email protected]. In addition, if you have further distributed this product, please identify your customers and notify them at once of this product recall. If you have any questions, please contact me at (224) 856-2250 or by e-mail, [email protected].

Distribution

US Distribution to state of MA.

Quantity

2,277 devices