Xhibit Central Station, Model 96102. Intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.

Recall: Z-0332-2017 · Initiated October 18, 2016

Recall

Recall Number
Z-0332-2017
Event Number
75477
Firm
Spacelabs Healthcare Inc
FEI Number
3010157426
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
October 18, 2016
Posted
November 2, 2016
Terminated
April 3, 2017
Address
35301 SE Center St, Snoqualmie, WA, 98065-9216

Description

Xhibit Central Station, Model 96102. Intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.

Reason

The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally.

Action

The firm sent the Urgent - Medical Device Correction letter, dated 18 October 2016, via priority service, return receipt requested, to all U.S. customers on 10/18/16. The firm will e-mail a customer letter (translated as necessary) to all international subsidiaries and distributors of record on October 25, 2016. Xhibit Central Stations used without Xhibit Telemetry Receivers, are not affected. Customers were asked to circulate the letter to all person affected and place a copy with the appropriate monitor user manual until corrections have been implemented. Spacelabs Healthcare will contact affected consignees to schedule a convenient time for Spacelabs to update the affected Xhibit Central Station(s) with a new software version 1.1.6 at no cost. Customers with questions can call Technical Support at 1-800-522-7025 and select 2.

Distribution

Worldwide Distribution -- USA, to AL, CA, CT, FL, GA, IA, ID, IL, IN, KY, MI, MN, MS, NC, NE, NJ, NY, OH, OK, PA, RI, SC, SD, TX, VA, WA, WY, and Puerto Rico; and, to the countries of AUSTRALIA, CANADA, CHILE, CZECH REPUBLIC, FRANCE, MEXICO, NETHERLANDS, PANAMA, SAUDI ARABIA, SWITZERLAND, TAIWAN, and UNITED KINGDOM.

Quantity

348 units (292 units in the US and 56 units outside the US)