FDA Recall Terminated

Excel Care Bariatric Bed, P 610A, Hill-Rom, Batesville, IN. Intended Use: The Excel Care Bariatric Bed is for bariatric patients with varying medical and physical conditions. The bed supports patients between 250 and 1000 pounds.

Recall: Z-0332-2011 · Initiated March 5, 2010

Recall

Recall Number
Z-0332-2011
Event Number
55301
Firm
Hill-Rom, Inc. 125 E Pearl St 1069 State Route 46 East Batesville IN 47006
FEI Number
1824206
Product Code
FNL
Status
Terminated
Root Cause
Device Design
Initiated
March 5, 2010
Posted
November 9, 2010
Terminated
October 6, 2014

Description

Excel Care Bariatric Bed, P 610A, Hill-Rom, Batesville, IN. Intended Use: The Excel Care Bariatric Bed is for bariatric patients with varying medical and physical conditions. The bed supports patients between 250 and 1000 pounds.

Reason

The firm received complaints regarding the siderails not operating correctly or becoming inoperable. The siderails could experience false latching, unintentional lowering or become inoperable. There have been reports of 5 patient falls and 2 injuries to caregivers.

Action

Hill-Rom sent out "Urgent Medical Device Correction" letters dated March 5, 2010. The letters described the product, problem and actions to be taken by the customers. The letters instructed the customers to locate the beds and examine the siderails for any visible damage (cracks, bending, loose parts) and conduct a physical assessment of all four siderails on each bed to determine if the sideral could be placed in all positions. If the siderails were found to be inoperable, the bed was to be taken out of service until a qualified Hill-Rom technician can make any necessary correction. The customers were instructed to contact Hill-Rom to make arrangements and inform them of all beds that they have loaned, sold, no longer use or disposed of at 800-445-3720, option 2. The letter states the firm is evaluating a design change to the siderail. If you have any questions concerning this request or the procedure as outlined, please contact Hill-Rom Technical Support at 800-445-3720, option 2.

Distribution

Worldwide distribution: USA states including: AK, FL, HI, IN, KS, LA, MS, OK, PA, TN, VT, and WA and countries including: Canada, Mexico, Angola, Australia, Saudi Arabia, Qatar, and United Emirates.

Quantity

1,406 beds both models.