FDA Recall Terminated

Thermo SCIENTIFIC, Sensititre CMC5VGNF, Gram Negative IVD AST

Recall: Z-0331-2023 · Initiated October 20, 2022

Recall

Recall Number
Z-0331-2023
Event Number
91104
Firm
Remel Inc
FEI Number
1924669
Product Code
JWY
Status
Terminated
Root Cause
Process control
Initiated
October 20, 2022
Posted
December 8, 2022
Terminated
September 30, 2024
Address
12076 Santa Fe Trail Dr, Lenexa, KS, 66215-3519

Description

Thermo SCIENTIFIC, Sensititre CMC5VGNF, Gram Negative IVD AST

Reason

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Action

The firm issued an "Urgent Medical Device Correction" notice to its consignees on 10/25/2022 by US mail. The notice explained the issue, risk to health, and notified the consignees of an update to the IFU, specifically, "to add limitations for Carbapenems (Imipenem, Doripenem, Meropenem, Ertapenem), Cefepime, Piperacillin/Tazobactam and Aztreonam added for users to "NOT REPORT" any result for each member of the Protea tribe (including Proteus spp., Providencia spp., Morganella Morganii) to address false susceptible results." Please complete and return the Required Customer Reply form via email to [email protected]. Please keep this notification on file. The notice needs to be passed on to all who need to be aware within your organization or to any organization where the potentially affected products have been transferred. This correction notification is being made and performed by the US agent, Remel Inc., on behalf of the Foreign Manufacturer of the affected product, Trek Diagnostic Systems Ltd., located in East Grinstead, UK. Trek Diagnostic Systems and Remel Inc., both part of Thermo Fisher Scientific, are investigating the root cause and will implement corrective actions to prevent future recurrence. If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or email [email protected].

Distribution

US Nationwide Distribution

Quantity

588 units