FindrWIRZ Guidewire System Catalog# 40-05 Cardiology: The system is intended for use in the cardiovascular system for introduction and positioning of over the-wire catheters and therapeutic devices during interventional procedures.
Recall
- Recall Number
- Z-0328-2017
- Event Number
- 75395
- Firm
- SentreHeart Inc
- FEI Number
- 3006482885
- Product Code
- DQX
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- September 29, 2016
- Posted
- November 9, 2016
- Terminated
- March 1, 2017
- Address
- 300 Saginaw Dr, Redwood City, CA, 94063-4743
Description
FindrWIRZ Guidewire System Catalog# 40-05 Cardiology: The system is intended for use in the cardiovascular system for introduction and positioning of over the-wire catheters and therapeutic devices during interventional procedures.
A complaint was received on September 27th, 2016 at a medical center in Germany for delamination and flaking of the PTFE lubricious coating on three FindrWIRZ Guidewire System .025 guidewires removed from their packaging hoops.
SentreHeart sent an Urgent Medical Device Voluntary Recall letter dated September 29, 2016, to all affected customers by traceable delivery method. Letters explained the issue and requested that any devices from the two affected lots be returned to SentreHEART immediately. Customers were asked to complete and return the enclosed response form as soon as possible via fax at 650-354-1204 or email to [email protected]. Customers with questions were instructed to call 1-855-256-7330. For questions regarding this recall call 650-241-6008.
Worldwide Distribution - US including MO, CA, OR, GA. NH, NY, MI, TX, MD, PA, IL, KS, NE, CT, AZ and Internationally to United Kingdom and Germany
132 worldwide distribution (98) of those units in US