VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Recall: Z-0327-2026
·
Initiated October 16, 2025
Recall
- Recall Number
- Z-0327-2026
- Event Number
- 97735
- Firm
- Canon Medical System, USA, INC.
- FEI Number
- 2020563
- Product Code
- LNH
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- October 16, 2025
- Posted
- October 24, 2025
- Address
- 2441 Michelle Dr, Tustin, CA, 92780-7047
Description
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Reason
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Action
Cannon Medical began notifying consignees via email on about 10/16/2025. Consignees were instructed that a representative of Canon Medical Systems USA will contact them to schedule a time to replace the relevant parts. If the message "Gradient power supply error is displayed before the replacement is performed, immediately stop using the system and contact your service representative. Consignees were requested to complete and return the provided response form
Distribution
US distribution to Ohio and Nevada
Quantity
2 units