NxStage System One S Cycler (High Permeability Hemodialysis System) Model no. NX1000-5-A
Recall
- Recall Number
- Z-0327-2016
- Event Number
- 72543
- Firm
- NxStage Medical, Inc.
- FEI Number
- 3003464075
- Product Code
- KDI
- Status
- Terminated
- Root Cause
- Software Design Change
- Initiated
- October 29, 2015
- Terminated
- April 29, 2019
- Address
- 350 Merrimack Street, Lawrence, MA, 01843-1748
Description
NxStage System One S Cycler (High Permeability Hemodialysis System) Model no. NX1000-5-A
Ultrafiltration (UF) Volume software error inaccurate fluid removal
NxStage issued issued a Field Safety Notice (FSN) on 10/29/15 to consignees who have NX1000-5 or a NX1000-5-A cycler. The FSN included Instructions for steps to be taken so that the cycler and treatments are not affected by the software error. NxStage will be sending a representative to the facility to attach detailed instructions on each cycler as a user reference. Return the attached reply form to acknowledge receipt of this letter. Questions or comments, please contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243).
Worldwide Distribution. US Nationwide, Netherlands, and Sweden.
16 units