FDA Recall Terminated

NxStage System One S Cycler (High Permeability Hemodialysis System) Model no. NX1000-5-A

Recall: Z-0327-2016 · Initiated October 29, 2015

Recall

Recall Number
Z-0327-2016
Event Number
72543
Firm
NxStage Medical, Inc.
FEI Number
3003464075
Product Code
KDI
Status
Terminated
Root Cause
Software Design Change
Initiated
October 29, 2015
Terminated
April 29, 2019
Address
350 Merrimack Street, Lawrence, MA, 01843-1748

Description

NxStage System One S Cycler (High Permeability Hemodialysis System) Model no. NX1000-5-A

Reason

Ultrafiltration (UF) Volume software error inaccurate fluid removal

Action

NxStage issued issued a Field Safety Notice (FSN) on 10/29/15 to consignees who have NX1000-5 or a NX1000-5-A cycler. The FSN included Instructions for steps to be taken so that the cycler and treatments are not affected by the software error. NxStage will be sending a representative to the facility to attach detailed instructions on each cycler as a user reference. Return the attached reply form to acknowledge receipt of this letter. Questions or comments, please contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243).

Distribution

Worldwide Distribution. US Nationwide, Netherlands, and Sweden.

Quantity

16 units