FDA Recall Terminated

VITROS Chemistry Products UPRO Slides packaged as 90 slides/pack catalog number 6800120 VITROS UPRO Slides quantitatively measure protein (UPRO) concentration in urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Recall: Z-0327-2014 · Initiated June 12, 2013

Recall

Recall Number
Z-0327-2014
Event Number
65572
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
JGP
Status
Terminated
Root Cause
Equipment maintenance
Initiated
June 12, 2013
Posted
November 15, 2013
Terminated
April 11, 2017
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS Chemistry Products UPRO Slides packaged as 90 slides/pack catalog number 6800120 VITROS UPRO Slides quantitatively measure protein (UPRO) concentration in urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason

Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.

Action

Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue. In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges. Complete and return the Confirmation of Receipt by June 21, 2013. If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311. For questions regarding this recall call 585-453-4224.

Distribution

Worldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY including Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, Germany, France, Italy, and Spain.

Quantity

31,395 units (10,864 units domestically & 20,531 units internationally)