24CM ALTA EXCHANGE KIT Model: AL19SH24-XP ALTA EXCHANGE KIT 15.5 Fr polyurethane twin lumen fixed tip hemodialysis catheter, cuff and side holes for chronic hemodialysis and apheresis. Latex free
Recall
- Recall Number
- Z-0327-2010
- Event Number
- 53605
- Firm
- Spire Biomedical, Inc. One Patriots Park
- FEI Number
- 1000159376
- Product Code
- MSD
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- October 23, 2009
- Posted
- November 17, 2009
- Terminated
- June 9, 2010
- Address
- Bedford, MA, 01730-2343
Description
24CM ALTA EXCHANGE KIT Model: AL19SH24-XP ALTA EXCHANGE KIT 15.5 Fr polyurethane twin lumen fixed tip hemodialysis catheter, cuff and side holes for chronic hemodialysis and apheresis. Latex free
There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira
Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.
Worldwide distribution - Nationwide, Government. Foreign: Canada, France.
26 units