DJO EMPOWR KNEE LOCKING TIBIAL PUNCH GUIDE, REF 801-05-042, NON STERILE, Qty 1 For use in orthopedic surgery.
Recall
- Recall Number
- Z-0326-2018
- Event Number
- 78727
- FEI Number
- 1000116912
- Product Code
- JWH
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- December 12, 2017
- Terminated
- June 13, 2022
- Address
- 9800 Metric Blvd, Encore Medical, Lp, Austin, TX, 78758-5445
Description
DJO EMPOWR KNEE LOCKING TIBIAL PUNCH GUIDE, REF 801-05-042, NON STERILE, Qty 1 For use in orthopedic surgery.
The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.
The field safety notice was disseminated by email on 12/12/2017. The notice stated the following: "DJO Sales Agents Action: Please review the bulletin that is provided with this notice. This bulletin contains new instructions on the use of the Locking Punch Guide and Punch. Please ensure your surgeons are familiar with this information. DJO Surgical then requires that you complete the attached acknowledgment and email it to [email protected]. Please have this form returned by December 15, 2017 Once redesigned guides are available, they will be exchanged through a separate field action."
One medical device distributor in California.
2 units