FDA Recall Terminated

Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.

Recall: Z-0326-2017 · Initiated August 9, 2016

Recall

Recall Number
Z-0326-2017
Event Number
75197
Firm
Hill-Rom, Inc.
FEI Number
1824206
Product Code
FSA
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
August 9, 2016
Posted
October 31, 2016
Terminated
March 24, 2017
Address
1069 State Road 46 E, Batesville, IN, 47006-7520

Description

Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.

Reason

Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.

Action

On 8/9/2016, URGENT: MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this Field Safety Correction, please contact Hill-Rom Technical Support at 1-800-445-3720, your distributor or your Hill-Rom representative.

Distribution

WA NC OH PA MI Australia Sweden

Quantity

188