FDA Recall Terminated

Ingenia, Intera, Achieva and Achieve dStream MR systems on Software versions R5.1.7 and R5.1.8 with the Mobi View software option. Diagnostic imaging system.

Recall: Z-0325-2016 · Initiated August 21, 2014

Recall

Recall Number
Z-0325-2016
Event Number
69127
Firm
Philips Medical Systems, Inc.
FEI Number
1218950
Product Code
LNH
Status
Terminated
Root Cause
Software Design Change
Initiated
August 21, 2014
Posted
November 23, 2015
Terminated
May 12, 2021
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Ingenia, Intera, Achieva and Achieve dStream MR systems on Software versions R5.1.7 and R5.1.8 with the Mobi View software option. Diagnostic imaging system.

Reason

When a fused series of a sagittal, coronal or radial multi station scan is generated in MobiView, the resultant image order is reversed. This occurs with software release R5.1.7/R5.1.8.

Action

Philips Healthcare sent an Customer Information Letter dated July 2014, on August 21, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips will issue a Field Change Order to install updated software in a Service Pack that makes the slice numbering consistent for fused and unfused images. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377

Distribution

Worldwide Distribution - USA (nationwide) and to the countries of: Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, Denmark Estonia, Ethiopia. France, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Korea, Republic of, Libya, Malaysia, Mexico, Nepal, Netherlands, New Zealand, Norway, Oman, Palestinian territory, Peru, Poland, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates and United Kingdom.

Quantity

307 units