FDA Recall Open, Classified

Endo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system

Recall: Z-0323-2026 · Initiated September 16, 2025

Recall

Recall Number
Z-0323-2026
Event Number
97715
Firm
Surgical Theater Inc
FEI Number
3010197287
Product Code
LLZ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
September 16, 2025
Posted
October 24, 2025
Address
23645 Mercantile Rd, Ste M, Beachwood, OH, 44122-5936

Description

Endo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system

Reason

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Action

Surgical Theater issued an URGENT: Medical Device Software correction notice dated September 17, 2025 to its consignees on 09/22/2025 via email. The notice explained the issue, risk to health, and provided the following: Customers can continue to use the software version(s) noted above until a software patch is made available by Surgical Theater. Until the software patch is installed, please follow these steps: 1. Read and share this notice with all relevant staff. 2. Continue to use the 3D Fusion functions to review the fusion results for all case files to verify alignment of secondary layer(s) to primary layer and to adjust fusion as necessary, and 3. Acknowledge your receipt of this notification by completing and returning the included acknowledgement form via email to [email protected]. A software patch is currently undergoing verification and should be available in the next 10 days. Surgical Theater has revised the 2D Image Fusion function to ensure the x, y and z axis rotation values are consistently applied when generating the aligned Axial, Sagittal and Coronal images. In addition, the rotate left/right buttons in all three planes were modified so the volume rotation stays consistent with the respective plane orientation. A Surgical Theater representative will contact your facility to schedule installation once the patch is available. For questions regarding this notification contact: Kevin Murrock, Sr. Director of Quality & Regulatory, Tel. 330-472-6520 (Monday through Friday, 9:00 AM to 5:00 PM, Eastern Time), email: [email protected]

Distribution

Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.

Quantity

19 units