Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph
Recall
- Recall Number
- Z-0322-2025
- Event Number
- 95559
- Firm
- Baxter Healthcare Corporation
- FEI Number
- 1417572
- Product Code
- DPS
- Status
- Open, Classified
- Root Cause
- Software design
- Initiated
- October 11, 2024
- Posted
- November 7, 2024
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 10/11/2024 via USPS First Class Mail. The notice explained the issue, potential hazard to the patient, and provided instructions for the to the system IT administrator to correct the issue which is already available in the current version of software.
Worldwide distribution.
399 units