VITROS Chemistry Products Mg Slides packaged as 300 slides/pack catalog number 8255093 and packaged as 90 slides/pack catalog number 1921204 VITROS Mg Slides quantitatively measure magnesium (Mg) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Recall
- Recall Number
- Z-0318-2014
- Event Number
- 65572
- Firm
- Ortho-Clinical Diagnostics
- FEI Number
- 1000136573
- Product Code
- JGJ
- Status
- Terminated
- Root Cause
- Equipment maintenance
- Initiated
- June 12, 2013
- Posted
- November 15, 2013
- Terminated
- April 11, 2017
- Address
- 100 Indigo Creek Dr, Rochester, NY, 14626-5101
Description
VITROS Chemistry Products Mg Slides packaged as 300 slides/pack catalog number 8255093 and packaged as 90 slides/pack catalog number 1921204 VITROS Mg Slides quantitatively measure magnesium (Mg) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue. In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges. Complete and return the Confirmation of Receipt by June 21, 2013. If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311. For questions regarding this recall call 585-453-4224.
Worldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY including Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, Germany, France, Italy, and Spain.
101,976 units (39,909 units domestically & 62,067 units internationally)