FDA Recall Open, Classified

The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.

Recall: Z-0317-2025 · Initiated September 6, 2024

Recall

Recall Number
Z-0317-2025
Event Number
95450
Firm
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION Nishinokyo Kuwabara Cho Nakagyo-Ku
FEI Number
3002808139
Product Code
JAA
Status
Open, Classified
Root Cause
Software change control
Initiated
September 6, 2024
Posted
November 7, 2024
Address
1 Kyoto Japan

Description

The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.

Reason

X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.

Action

On 9/13/24, recall notices were emailed to consignees asking them to do the following: 1) If the problem occurs it is possible to solve this problem by turning off the power of the system and then pushing the reset switch on the Digital radiography unit s control cabinet. 2) Your firm authorized service representative will be contacting to schedule the implementation of the corrective action but if you are not contacted, contact the firm's National Support. 3) Complete and return the response form via email to response to [email protected] If you have questions contact the firm's National Technical Support at 844-487-2767, available 8:00AM-5:00PM, Monday - Friday.

Distribution

US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.

Quantity

30