FDA Recall Terminated

Periarticular locking plate elbow-sterile, Item Nos. 47235800503 47235800509 47235800511 47235800515 47235800519 47235800603 47235800605 47235800607 47235800609 47235800611 47235800615 47235800619 47235800705 47235800711 47235800715 47235800803 47235800805 47235800807 47235800811 47235800815 47235800911 47235800913 47235800915 47235801209 47235801211 47235801213 47235801215 47235802305 47235802307 47235802311 47235802315 47235802405 47235802407 47235802409 47235802411 47235802415 47235810707 47235810711 47235810715 47235810805 47235810807 47235810811 47235810815

Recall: Z-0317-2020 · Initiated October 10, 2019

Research recall Z-0317-2020 in seconds

The Research Assistant summarizes recall history, root causes and affected devices — and cites every record.

Free to try · every answer cites its records

Recall

Recall Number
Z-0317-2020
Event Number
83962
Firm
Zimmer Biomet, Inc.
FEI Number
1000220733
Product Code
HRS
Status
Terminated
Root Cause
Packaging change control
Initiated
October 10, 2019
Terminated
May 19, 2020
Address
1800 W Center St, Warsaw, IN, 46580-2304

Description

Periarticular locking plate elbow-sterile, Item Nos. 47235800503 47235800509 47235800511 47235800515 47235800519 47235800603 47235800605 47235800607 47235800609 47235800611 47235800615 47235800619 47235800705 47235800711 47235800715 47235800803 47235800805 47235800807 47235800811 47235800815 47235800911 47235800913 47235800915 47235801209 47235801211 47235801213 47235801215 47235802305 47235802307 47235802311 47235802315 47235802405 47235802407 47235802409 47235802411 47235802415 47235810707 47235810711 47235810715 47235810805 47235810807 47235810811 47235810815

Reason

Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).

Action

The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, as per standard orthopedic procedure planning, to reduce the likelihood of an extension of surgery time. 3. Examine devices thoroughly for adherent LDPE bag residue prior to use. Use a backup device if LDPE bag residue is detected. 4. Return any devices which display adherent LDPE bag residue to Zimmer Biomet, consistent with the standard instructions for use regarding the disposition of damaged components. 5. Complete the Certificate of Acknowledgement and return to the firm. 6. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facility's documentation. If you have further questions or concerns regarding this recall, please call the firm's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].

Distribution

Distributed nationwide.

Quantity

5,459,583 total devices