FDA Recall Terminated

FC 500 MPL Flow Cytometry System with MXP Software Version 2.0 & 2.1, Part Number 626554 FC 500 MPL 733313 MPL Hardware Upgrade Kit

Recall: Z-0317-2007 · Initiated October 17, 2006

Recall

Recall Number
Z-0317-2007
Event Number
36822
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
GKZ
Status
Terminated
Root Cause
Other
Initiated
October 17, 2006
Posted
December 27, 2006
Terminated
February 29, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

FC 500 MPL Flow Cytometry System with MXP Software Version 2.0 & 2.1, Part Number 626554 FC 500 MPL 733313 MPL Hardware Upgrade Kit

Reason

The ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialog box is accessed. Issue limited only to protocols utilizing the Ratio parameter such as LeukoSure and DNA Cell Cycle.

Action

A Product Corrective Action (PCA) letter was mailed on Oct 17, 2006 to FC500 MPL and MXP software accounts. They are informed that if the operator opens the Parameter Selection Dialogue box after parameters have been selected for the RATIO Parameter, the selected parameters could revert back to the default choices. If these same parameters were used in the creation of the RATIO Parameter in any other saved protocol, the protocols will also be changed once they are opened after the error occurs. An immediate user preventive action to remedy this issue each time acquisition of samples that require a protocol utilizing the ratio parameter is desired is given within the letter. Consignees are asked to complete and return the enclosed Response Form within 10 days.

Distribution

Nationwide and Canada

Quantity

37 units in the US; 4 units in Canada