FDA Recall Terminated

Distal Centralizer 10 mm

Recall: Z-0315-2011 · Initiated July 28, 2010

Recall

Recall Number
Z-0315-2011
Event Number
56499
Firm
Ortho Development Corporation
FEI Number
3017298928
Product Code
KWY
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 28, 2010
Posted
November 3, 2010
Terminated
November 4, 2010
Address
12187 South Business Park Dr, Draper, UT, 84020

Description

Distal Centralizer 10 mm

Reason

The hip stem did not fit the distal centralizer because the hole was not drilled with the appropriate taper.

Action

Ortho Development sent a Field Notification Memo dated July 28, 2010 to Japan Medical Dynamic Marketing instructing them to reconcile parts that may have been used and return unused parts to Ortho. Customers can contact Ortho Development at 1-801-619-3419.

Distribution

Worldwide Distribution: Japan

Quantity

39 units