FDA Recall Terminated

SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood

Recall: Z-0314-2017 · Initiated June 1, 2016

Recall

Recall Number
Z-0314-2017
Event Number
74786
Firm
BBI SOLUTIONS OEM LTD 8 Tom McDonald Avenue Dundee United Kingdom
FEI Number
3012801636
Product Code
GHH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
June 1, 2016
Terminated
June 14, 2018

Description

SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood

Reason

A component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.

Action

BBI Solutions sent an Urgent Medical Device Recall letter dated June 30, 2016, to all affected customers. Customers were advised to immediately discontinue use of the product, consider the requirement for review of reported test results, complete and return the customer acknowledgement form to BBI Solutions and to complete and return the customer reconciliation form to the distributor. Customers who have forwarded product to other sites, should inform those customers of the product recall and provide them with a copy of the notice. Customers with questions were instructed to call Tel: +44(0)1382 564093.

Distribution

US distribution to CT and IN

Quantity

420 units (183 units quarantined)