FDA Recall Terminated

COULTER LH 500 Series System, Part Number: 178833, Beckman Coulter, Brea, CA. (Hematology analyzer)

Recall: Z-0314-2008 · Initiated April 18, 2007

Recall

Recall Number
Z-0314-2008
Event Number
44916
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
GKZ
Status
Terminated
Root Cause
Software design
Initiated
April 18, 2007
Posted
December 22, 2007
Terminated
April 16, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

COULTER LH 500 Series System, Part Number: 178833, Beckman Coulter, Brea, CA. (Hematology analyzer)

Reason

Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2)- When manually entering a Patient ID, if a blank space is entered between th

Action

Product Corrective Action (PCA) letters were mailed on April 18,2007, to all GENS, LH500, LH750 and LH780 accounts. They are informed that a potential misidentification can occur in 2 scenarios. 1. When a positive identifier (Sample ID or CassetteIPosition) is manually edited to a positive identifier that is already in the ToDo list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2. When manually entering a Patient ID, if a blank space is entered between the characters in the Patient ID field, the system will only accept the characters before the space.----Short term fix----- The letter explains the issues and provides the customer work around. (Letters were sent by US mail). A customer response form was provided. The root cause of the problem is still under investigation, software upgrades, when approved will be made to devices.

Distribution

World wide: USA and Canada

Quantity

917 units in the US; 82 units to Canada