FDA Recall Terminated

PhoenixSpec Calibrator & Kit is used to calibrate the PhoenixSpec & Phoenix AP Instruments. The instruments are then used to measure the turbidity of microbial suspensions equivalent to McFarland standards 0.10 to 4.50.

Recall: Z-0311-2017 · Initiated June 1, 2016

Recall

Recall Number
Z-0311-2017
Event Number
75375
Firm
Becton Dickinson & Co. 7 Loveton Cir BD Diagnostic Systems
FEI Number
1119779
Product Code
LON
Status
Terminated
Root Cause
Process control
Initiated
June 1, 2016
Terminated
November 8, 2016
Address
Sparks, MD, 21152-9212

Description

PhoenixSpec Calibrator & Kit is used to calibrate the PhoenixSpec & Phoenix AP Instruments. The instruments are then used to measure the turbidity of microbial suspensions equivalent to McFarland standards 0.10 to 4.50.

Reason

BD has determined that the BBL PhoenixSpec Calibrator kits referenced in this recall from the following lots (5364776, 5345796 ) are labeled with an incorrect expiration date (year). The expiration date printed on the individual calibration tubes within the kit are correct but difficult to read on a portion of the tubes.

Action

All US customers were contacted via letters sent via UPS. Ex-US BD sites will contact customers in their local regions. The firm will use an acknowledgement form included with the follow up letter that is mailed or faxed back to BD will be reconciled with the customer list. BD will contact 100% of our customers that have not responded to the communication to verify acknowledgement

Distribution

BD has contacted all 31 US customers and 22 Ex-US sites which received product from the associated catalog numbers. US: AL, CA, CO, KY, MD, MI, MS, NE, NJ, OH, PA, SC, SD, TX, VA, WA and WI. Ex-US: Argentina, Europe, Canada, China, Ecuador, India, Japan, New Zealand, Peru, Singapore, Thailand and Vietnam 4 US government sites were contacted

Quantity

20