FDA Recall Terminated

Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled selective infusion of physician-specified therapeutics agents such as embolization materials and of diagnostic materials such as contrast media. UltraFlow HPC Flow: The device is intended for the controlled selective infusion of physician-specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuroanatomy. It is not intended for use in the coronary vasculature. Catalog No. 105-5055V02 105-5055V02 105-5055V05 105-5065V02 105-5065V05 105-5066V02 105-5066V05

Recall: Z-0309-2017 · Initiated October 5, 2016

Recall

Recall Number
Z-0309-2017
Event Number
75358
Firm
Micro Therapeutics Inc, Dba Ev3 Neurovascular
FEI Number
3004904811
Product Code
KRA
Status
Terminated
Root Cause
Device Design
Initiated
October 5, 2016
Posted
November 9, 2016
Terminated
October 11, 2017
Address
9775 Toledo Way, Irvine, CA, 92618-1811

Description

Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled selective infusion of physician-specified therapeutics agents such as embolization materials and of diagnostic materials such as contrast media. UltraFlow HPC Flow: The device is intended for the controlled selective infusion of physician-specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuroanatomy. It is not intended for use in the coronary vasculature. Catalog No. 105-5055V02 105-5055V02 105-5055V05 105-5065V02 105-5065V05 105-5066V02 105-5066V05

Reason

Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.

Action

Medtronic sent an Urgent Medical Device Recall letter dated October 5, 2016, to all affected customers to inform them that Medtronic has identified the potential for an issue with a specific set of lot numbers of Medtronic Neurovascular products as noted where the PTFE (polytetrafluoroethylene) coating could delaminate and separate from the delivery wire. Customers are instructed to complete the attached customer confirmation certificate and fax it to Medtronic at (949) 434-5020 to the attention of Neurovascular Quality. On October 14, 2016, Medtronic announced that it has notified customers of a voluntary recall of certain lots of its Pipeline embolization device, Alligator retrieval device and X-Celerator hydrophilic guidewire. The recall also includes the stylet containing UltraFlow flow directed micro catheters and Marathon flow directed micro catheters. This voluntary recall is being conducted due to the potential separation and detachment of the polytetrafluoroethylene (PTFE) coating on parts of these devices. Should the PTFE separate from the delivery wire or stylets, PTFE particulate could enter the blood stream of the patient. PTFE in the blood stream, based on the size and quantity, could lead to a thromboembolic event. Customers were asked to complete the attached Customer Confirmation Certificate and fax it to Medtronic at 1-949-434-5020 to the attention of Neurovascular Quality. Customers with questions were instructed to contact their Medtronic representative. On 1/12/17, an Urgent Medical Device Recall letter was sent to expand the recall to include additional lots.

Distribution

Worldwide Distribution - US (nationwide) and Internationally to Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Japan, Korea, Latvia, Lithuania, Malaysia, Mexico, Nepal, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Russian Federation, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.

Quantity

66,429 units