FDA Recall Terminated

Titanium Orthos Bracket

Recall: Z-0309-05 · Initiated November 18, 2004

Recall

Recall Number
Z-0309-05
Event Number
30515
Firm
Ormco Corporation
FEI Number
2016150
Product Code
EJF
Status
Terminated
Root Cause
Other
Initiated
November 18, 2004
Posted
December 15, 2004
Terminated
June 17, 2005
Address
1332 S Lone Hill Ave, Glendora, CA, 91740-5339

Description

Titanium Orthos Bracket

Reason

Misassembly. Stainless steel and titanium parts were mistakenly welded together to construct the device. The weld failed.

Action

Recall letters were mailed from 11/18 thru 11/23. Letters request return of product.

Distribution

Nationwide and France, Luxembourg, Spain, Netherlands, Russia, Ireland, Italy, Germany, United Kingdom, Hungary, Finland, Denmark.

Quantity

6,257