Pipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracranial aneurysms (lAs) in the internal carotid artery from the petrous to the superior hypophyseal segements. Catalog No. FA-71300-25V09 FA-71350-25V16 FA-71400-25V10 FA-71425-35V13 FA-71475-30V16 FA-71300-25V12 FA-71350-25V17 FA-71400-25V11 FA-71425-35V15 FA-71475-30V18 FA-71300-25V13 FA-71350-25V18 FA-71400-25V12 FA-71425-35V18 FA-71475-35V09 FA-71300-25V15 FA-71350-25V19 FA-71400-25V13 FA-71450-25V09 FA-71475-35V10 FA-71300-25V16 FA-71350-30V09 FA-71400-25V15 FA-71450-25V10 FA-71475-35V11 FA-71300-25V17 FA-71350-30V10 FA-71400-25V16 FA-71450-25V11 FA-71475-35V12 FA-71300-25V18 FA-71350-30V12 FA-71400-25V18 FA-71450-25V12 FA-71475-35V13 FA-71300-25V19 FA-71350-30V13 FA-71400-25V19 FA-71450-25V13 FA-71475-35V15 FA-71300-30V09 FA-71350-30V15 FA-71400-30V09 FA-71450-25V15 FA-71475-35V18 FA-71300-30V12 FA-71350-30V16 FA-71400-30V10 FA-71450-25V16 FA-71475-35V19 FA-71300-30V13 FA-71350-30V18 FA-71400-30V11 FA-71450-25V18 FA-71500-25V09 FA-71300-30V15 FA-71350-35V09 FA-71400-30V12 FA-71450-25V19 FA-71500-25V10 FA-71300-30V16 FA-71350-35V10 FA-71400-30V13 FA-71450-30V09 FA-71500-25V11 FA-71300-30V19 FA-71350-35V12 FA-71400-30V15 FA-71450-30V10 FA-71500-25V12 FA-71300-35V09 FA-71350-35V13 FA-71400-30V16 FA-71450-30V11 FA-71500-25V13 FA-71300-35V12 FA-71350-35V15 FA-71400-30V18 FA-71450-30V12 FA-71500-25V15 FA-71300-35V13 FA-71350-35V18 FA-71400-35V09 FA-71450-30V13 FA-71500-25V16 FA-71300-35V15 FA-71375-25V09 FA-71400-35V13 FA-71450-30V15 FA-71500-25V18 FA-71325-25V09 FA-71375-25V10 FA-71400-35V15 FA-71450-30V16 FA-71500-25V19 FA-71325-25V10 FA-71375-25V11 FA-71400-35V18 FA-71450-30V18 FA-71500-30V09 FA-71325-25V12 FA-71375-25V12 FA-71425-25V09 FA-71450-30V19 FA-71500-30V10 FA-71325-25V13 FA-71375-25V13 FA-71425-25V10 FA-71450-35V09 FA-71500-30V11 FA-71325-25V15 FA-71375-25V15 FA-71425-25V11 FA-71450-35V10 FA-71500-30V12 FA-71325-25V18 FA-71375-25V16 FA-71425-25V12 FA-71450-35V12 FA-71500-30V13 FA-71325-25V19 FA-71375-25V17 FA-71425-25V13 FA-71450-35V13 FA-71500-30V15 FA-71325-30V09 FA-71375-25V18 FA-71425-25V15 FA-71450-35V15 FA-71500-30V16 FA-71325-30V10 FA-71375-25V19 FA-71425-25V16 FA-71450-35V16 FA-71500-30V17 FA-71325-30V12 FA-71375-30V09 FA-71425-25V18 FA-71475-25V09 FA-71500-30V18 FA-71325-30V13 FA-71375-30V10 FA-71425-25V19 FA-71475-25V10 FA-71500-35V09 FA-71325-30V15 FA-71375-30V12 FA-71425-30V09 FA-71475-25V11 FA-71500-35V10 FA-71325-30V16 FA-71375-30V13 FA-71425-30V10 FA-71475-25V12 FA-71500-35V11 FA-71325-30V18 FA-71375-30V15 FA-71425-30V11 FA-71475-25V13 FA-71500-35V12 FA-71325-30V19 FA-71375-30V16 FA-71425-30V12 FA-71475-25V15 FA-71500-35V13 FA-71325-35V09 FA-71375-30V18 FA-71425-30V13 FA-71475-25V16 FA-71500-35V15 FA-71325-35V12 FA-71325-35V13 FA-71325-35V15 FA-71350-25V09 FA-71350-25V10 FA-71350-25V13 FA-71350-25V15 FA-77250-10V13 FA-77250-10V19 FA-77250-12V11 FA-77250-12V12 FA-77250-12V13 FA-77250-12V18 FA-77250-14V12 FA-77250-14V13 FA-77250-14V15 FA-77250-14V18 FA-77250-14V19 FA-77250-16V11 FA-77250-16V12 FA-77250-16V13 FA-77250-16V15 FA-77250-16V16 FA-77250-16V19 FA-77250-18V09 FA-77250-18V10 FA-77250-18V11 FA-77250-18V12 FA-77250-18V13 FA-77250-18V16 FA-77250-18V18 FA-77250-18V19 FA-77250-20V10 FA-77250-20V11 FA-71375-35V09 FA-71375-35V12 FA-71375-35V13 FA-71375-35V15 FA-71375-35V16 FA-71375-35V18 FA-71400-25V09 FA-77275-14V11 FA-77275-14V12 FA-77275-14V13 FA-77275-14V17 FA-77275-14V18 FA-77275-14V19 FA-77275-16V12 FA-77275-16V13 FA-77275-16V16 FA-77275-16V18 FA-77275-16V19 FA-772
Recall
- Recall Number
- Z-0307-2017
- Event Number
- 75358
- Firm
- Micro Therapeutics Inc, Dba Ev3 Neurovascular
- FEI Number
- 3004904811
- Product Code
- OUT
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 5, 2016
- Posted
- November 9, 2016
- Terminated
- October 11, 2017
- Address
- 9775 Toledo Way, Irvine, CA, 92618-1811
Description
Pipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracranial aneurysms (lAs) in the internal carotid artery from the petrous to the superior hypophyseal segements. Catalog No. FA-71300-25V09 FA-71350-25V16 FA-71400-25V10 FA-71425-35V13 FA-71475-30V16 FA-71300-25V12 FA-71350-25V17 FA-71400-25V11 FA-71425-35V15 FA-71475-30V18 FA-71300-25V13 FA-71350-25V18 FA-71400-25V12 FA-71425-35V18 FA-71475-35V09 FA-71300-25V15 FA-71350-25V19 FA-71400-25V13 FA-71450-25V09 FA-71475-35V10 FA-71300-25V16 FA-71350-30V09 FA-71400-25V15 FA-71450-25V10 FA-71475-35V11 FA-71300-25V17 FA-71350-30V10 FA-71400-25V16 FA-71450-25V11 FA-71475-35V12 FA-71300-25V18 FA-71350-30V12 FA-71400-25V18 FA-71450-25V12 FA-71475-35V13 FA-71300-25V19 FA-71350-30V13 FA-71400-25V19 FA-71450-25V13 FA-71475-35V15 FA-71300-30V09 FA-71350-30V15 FA-71400-30V09 FA-71450-25V15 FA-71475-35V18 FA-71300-30V12 FA-71350-30V16 FA-71400-30V10 FA-71450-25V16 FA-71475-35V19 FA-71300-30V13 FA-71350-30V18 FA-71400-30V11 FA-71450-25V18 FA-71500-25V09 FA-71300-30V15 FA-71350-35V09 FA-71400-30V12 FA-71450-25V19 FA-71500-25V10 FA-71300-30V16 FA-71350-35V10 FA-71400-30V13 FA-71450-30V09 FA-71500-25V11 FA-71300-30V19 FA-71350-35V12 FA-71400-30V15 FA-71450-30V10 FA-71500-25V12 FA-71300-35V09 FA-71350-35V13 FA-71400-30V16 FA-71450-30V11 FA-71500-25V13 FA-71300-35V12 FA-71350-35V15 FA-71400-30V18 FA-71450-30V12 FA-71500-25V15 FA-71300-35V13 FA-71350-35V18 FA-71400-35V09 FA-71450-30V13 FA-71500-25V16 FA-71300-35V15 FA-71375-25V09 FA-71400-35V13 FA-71450-30V15 FA-71500-25V18 FA-71325-25V09 FA-71375-25V10 FA-71400-35V15 FA-71450-30V16 FA-71500-25V19 FA-71325-25V10 FA-71375-25V11 FA-71400-35V18 FA-71450-30V18 FA-71500-30V09 FA-71325-25V12 FA-71375-25V12 FA-71425-25V09 FA-71450-30V19 FA-71500-30V10 FA-71325-25V13 FA-71375-25V13 FA-71425-25V10 FA-71450-35V09 FA-71500-30V11 FA-71325-25V15 FA-71375-25V15 FA-71425-25V11 FA-71450-35V10 FA-71500-30V12 FA-71325-25V18 FA-71375-25V16 FA-71425-25V12 FA-71450-35V12 FA-71500-30V13 FA-71325-25V19 FA-71375-25V17 FA-71425-25V13 FA-71450-35V13 FA-71500-30V15 FA-71325-30V09 FA-71375-25V18 FA-71425-25V15 FA-71450-35V15 FA-71500-30V16 FA-71325-30V10 FA-71375-25V19 FA-71425-25V16 FA-71450-35V16 FA-71500-30V17 FA-71325-30V12 FA-71375-30V09 FA-71425-25V18 FA-71475-25V09 FA-71500-30V18 FA-71325-30V13 FA-71375-30V10 FA-71425-25V19 FA-71475-25V10 FA-71500-35V09 FA-71325-30V15 FA-71375-30V12 FA-71425-30V09 FA-71475-25V11 FA-71500-35V10 FA-71325-30V16 FA-71375-30V13 FA-71425-30V10 FA-71475-25V12 FA-71500-35V11 FA-71325-30V18 FA-71375-30V15 FA-71425-30V11 FA-71475-25V13 FA-71500-35V12 FA-71325-30V19 FA-71375-30V16 FA-71425-30V12 FA-71475-25V15 FA-71500-35V13 FA-71325-35V09 FA-71375-30V18 FA-71425-30V13 FA-71475-25V16 FA-71500-35V15 FA-71325-35V12 FA-71325-35V13 FA-71325-35V15 FA-71350-25V09 FA-71350-25V10 FA-71350-25V13 FA-71350-25V15 FA-77250-10V13 FA-77250-10V19 FA-77250-12V11 FA-77250-12V12 FA-77250-12V13 FA-77250-12V18 FA-77250-14V12 FA-77250-14V13 FA-77250-14V15 FA-77250-14V18 FA-77250-14V19 FA-77250-16V11 FA-77250-16V12 FA-77250-16V13 FA-77250-16V15 FA-77250-16V16 FA-77250-16V19 FA-77250-18V09 FA-77250-18V10 FA-77250-18V11 FA-77250-18V12 FA-77250-18V13 FA-77250-18V16 FA-77250-18V18 FA-77250-18V19 FA-77250-20V10 FA-77250-20V11 FA-71375-35V09 FA-71375-35V12 FA-71375-35V13 FA-71375-35V15 FA-71375-35V16 FA-71375-35V18 FA-71400-25V09 FA-77275-14V11 FA-77275-14V12 FA-77275-14V13 FA-77275-14V17 FA-77275-14V18 FA-77275-14V19 FA-77275-16V12 FA-77275-16V13 FA-77275-16V16 FA-77275-16V18 FA-77275-16V19 FA-772
Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.
Medtronic sent an Urgent Medical Device Recall letter dated October 5, 2016, to all affected customers to inform them that Medtronic has identified the potential for an issue with a specific set of lot numbers of Medtronic Neurovascular products as noted where the PTFE (polytetrafluoroethylene) coating could delaminate and separate from the delivery wire. Customers are instructed to complete the attached customer confirmation certificate and fax it to Medtronic at (949) 434-5020 to the attention of Neurovascular Quality. On October 14, 2016, Medtronic announced that it has notified customers of a voluntary recall of certain lots of its Pipeline embolization device, Alligator retrieval device and X-Celerator hydrophilic guidewire. The recall also includes the stylet containing UltraFlow flow directed micro catheters and Marathon flow directed micro catheters. This voluntary recall is being conducted due to the potential separation and detachment of the polytetrafluoroethylene (PTFE) coating on parts of these devices. Should the PTFE separate from the delivery wire or stylets, PTFE particulate could enter the blood stream of the patient. PTFE in the blood stream, based on the size and quantity, could lead to a thromboembolic event. Customers were asked to complete the attached Customer Confirmation Certificate and fax it to Medtronic at 1-949-434-5020 to the attention of Neurovascular Quality. Customers with questions were instructed to contact their Medtronic representative. On 1/12/17, an Urgent Medical Device Recall letter was sent to expand the recall to include additional lots.
Worldwide Distribution - US (nationwide) and Internationally to Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Japan, Korea, Latvia, Lithuania, Malaysia, Mexico, Nepal, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Russian Federation, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
11,860 units