FDA Recall Terminated

Giraffe OmniBed, labeled in part: ***GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com***. The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand.

Recall: Z-0305-2009 · Initiated June 20, 2008

Recall

Recall Number
Z-0305-2009
Event Number
49097
Firm
Datex Ohmeda, Inc. dba GE Healthcare
FEI Number
1121732
Product Code
FMZ
Status
Terminated
Root Cause
Equipment maintenance
Initiated
June 20, 2008
Posted
November 21, 2008
Terminated
March 24, 2010
Address
8880 Gorman Rd, Laurel, MD, 20723-5800

Description

Giraffe OmniBed, labeled in part: ***GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com***. The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand.

Reason

Manufacturer of neonatal incubators changed the device labeling to account for new use and care instructions.

Action

A notification letter (Urgent Medical Device Correction) was sent to consignees on 06/20/08. The notification letter addressed healthcare professionals, including neonatal, infection control and nursing departments at medical facilities. In add tion, the letter advised of revisions to the current Giraffe OmniBed cleaning instructions contained in the product Operators Manual. Instructions were enclosed with the notification letter and are intended to supplement those already present in the manual. Contact the GE Healthcare Technical Support center at 1-800-558-7044 for questions or additional information. If outside of the USA, please contact your local GE Healthcare Service Representative.

Distribution

Worldwide Distribution: The devices were widely distributed to national and international medical facilities.

Quantity

7,878 beds