FDA Recall Terminated

Toshiba Aplio XG Diagnostic Ultrasound System, Model: SSA-790A with version 1.1 software, Toshiba America Medical Systems, Inc., Tustin, CA. 92780

Recall: Z-0305-2008 · Initiated October 2, 2007

Recall

Recall Number
Z-0305-2008
Event Number
45811
Firm
Toshiba American Med Sys Inc
FEI Number
2020563
Product Code
IYN
Status
Terminated
Root Cause
Software design
Initiated
October 2, 2007
Posted
January 9, 2008
Terminated
April 20, 2012
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92781-2068

Description

Toshiba Aplio XG Diagnostic Ultrasound System, Model: SSA-790A with version 1.1 software, Toshiba America Medical Systems, Inc., Tustin, CA. 92780

Reason

Panel Lock-up: When using dynamic image acquisition modes the panel may lock up, if the operator attempts to perform a still image capture. If the panel locks up the system will have to be rebooted to continue the examination which may cause a delay in completion of the patient examination.

Action

Notification letters, Urgent Medical Device Correction, with an attached Customer Response form were sent to customers beginning on October 2, 2007. Customers were notified this action was being taken as a result of an investigation that resulted from a software issue that was found by the manufacturer. Customers were informed that a Toshiba Customer Service Engineer would contact them to schedule a software upgrade when the software becomes available. If customers have any questions they were instructed to contact their or call (800) 421-1968 and ask for the Regulatory Affairs Department.--A short term instruction included "Avoid using the Freeze Key when performing dynamic image acquisitions"

Distribution

Nationwide: USA including the states of AK, AZ, CA, FL, IA, IL, IN, KS, MI, MN, MO, MT, NE, NV, NY, OH, OR, PR, VA, WI & WV

Quantity

69