FDA Recall Terminated

Histochemical PAS (Periodic Acid Schiff) Reaction Set. in vitro diagnostic. Product size configuration: Set of 4 bottles including Schiff Reagent 225mL, Light Green SF Yellowish Stain 225mL, Sodium Carbonate Solution 225mL, and Periodic Acid Solution 225mL. Each kit provides reagents sufficient for a maximum of 100 tests.

Recall: Z-0305-2007 · Initiated July 25, 2006

Recall

Recall Number
Z-0305-2007
Event Number
36096
Firm
EMD Chemicals Inc
FEI Number
2244894
Product Code
KKS
Status
Terminated
Root Cause
Other
Initiated
July 25, 2006
Posted
December 27, 2006
Terminated
October 6, 2008
Address
480 S Democrat Rd, Gibbstown, NJ, 08027-1239

Description

Histochemical PAS (Periodic Acid Schiff) Reaction Set. in vitro diagnostic. Product size configuration: Set of 4 bottles including Schiff Reagent 225mL, Light Green SF Yellowish Stain 225mL, Sodium Carbonate Solution 225mL, and Periodic Acid Solution 225mL. Each kit provides reagents sufficient for a maximum of 100 tests.

Reason

Due to loose caps, the SO2 levels dropped significantly from release levels causing the product not to perform properly.

Action

On July 27, 2006 EMD contacted its direct account customers via letter. The firm requested discontinue use of the product and contact EMD for return.

Distribution

Nationwide.

Quantity

74 kits