FDA Recall
Terminated
Infinity Docking Station (IDS) (transportable patient monitor) Repair Kit, part number 7262814
Recall: Z-0303-2007
·
Initiated November 17, 2006
Recall
- Recall Number
- Z-0303-2007
- Event Number
- 36856
- Firm
- Draeger Medical, Inc.
- FEI Number
- 2517967
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 17, 2006
- Posted
- December 19, 2006
- Terminated
- November 19, 2008
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Infinity Docking Station (IDS) (transportable patient monitor) Repair Kit, part number 7262814
Reason
Top cover of assembly may separate. The situation can result in the monitor falling, which could result in an injury to the clinician or patient.
Action
The recaling firm issued a recall letter to the US customers 11/17/2006 via certified mail. An authorized service representative will inspect each potentially affected unit. If a unit is found to have an affected docking station, the unit will be repaired or replaced free or charge. Notification to international customers will occur through the international subsidiaries.
Distribution
Nationwide
Quantity
23,345 units - 4336 US