FDA Recall Terminated

FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip (SPSA-20002590) Provide fractional treatments in gynecology.

Recall: Z-0301-2017 · Initiated September 25, 2016

Recall

Recall Number
Z-0301-2017
Event Number
75384
Firm
Lumenis Limited 13 Hayetzira St., Yokneam Ind. Park Yokneam Israel
FEI Number
3004135191
Product Code
GEX
Status
Terminated
Root Cause
Software Manufacturing/Software Deployment
Initiated
September 25, 2016
Posted
October 23, 2016
Terminated
April 11, 2017

Description

FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip (SPSA-20002590) Provide fractional treatments in gynecology.

Reason

The cleaning instructions provided in FemTouch Operator's Manual might be ineffective for a fully assembled device and may lead to a potential for cross-contamination.

Action

On 9/25/16 consignees were notified via an Urgent - Field Safety Corrective Action letter with a FemTouch disassembly tool kit as well as FemTouch disassembly instructions. The instructions describe the simple disassembly process to be performed as part of the Pre-cleaning phase and the process for reassembling the device prior to re-use. It is recommended that customers clean and sterilize all of their handpieces according to the pre-cleaning disassembly process before each use. Consignees were asked to complete and return a verification form. Questions about the FSCA, should be directed to the Lumenis field correction administrator, Rafi Barel at [email protected].

Distribution

Worldwide Distribution -- USA, Italy, Russian Fed., Australia, United Arab Emirates, Spain, United Kingdom, India, Kuwait, and Belarus.

Quantity

69 units