FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip (SPSA-20002590) Provide fractional treatments in gynecology.
Recall
- Recall Number
- Z-0301-2017
- Event Number
- 75384
- Firm
- Lumenis Limited 13 Hayetzira St., Yokneam Ind. Park Yokneam Israel
- FEI Number
- 3004135191
- Product Code
- GEX
- Status
- Terminated
- Root Cause
- Software Manufacturing/Software Deployment
- Initiated
- September 25, 2016
- Posted
- October 23, 2016
- Terminated
- April 11, 2017
Description
FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip (SPSA-20002590) Provide fractional treatments in gynecology.
The cleaning instructions provided in FemTouch Operator's Manual might be ineffective for a fully assembled device and may lead to a potential for cross-contamination.
On 9/25/16 consignees were notified via an Urgent - Field Safety Corrective Action letter with a FemTouch disassembly tool kit as well as FemTouch disassembly instructions. The instructions describe the simple disassembly process to be performed as part of the Pre-cleaning phase and the process for reassembling the device prior to re-use. It is recommended that customers clean and sterilize all of their handpieces according to the pre-cleaning disassembly process before each use. Consignees were asked to complete and return a verification form. Questions about the FSCA, should be directed to the Lumenis field correction administrator, Rafi Barel at [email protected].
Worldwide Distribution -- USA, Italy, Russian Fed., Australia, United Arab Emirates, Spain, United Kingdom, India, Kuwait, and Belarus.
69 units