FDA Recall Terminated

Baxter CAPD Solution Transfer Set for use with UV-Flash Germicidal Exchange Device, 1.2 m (48''); a sterile set consisting of a bag connector (spike connector), on-off clamp assembly, tubing and double sealing male Luer lock connector; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085; product code R5C4325

Recall: Z-0301-05 · Initiated November 5, 2004

Recall

Recall Number
Z-0301-05
Event Number
30495
Firm
Baxter Healthcare Renal Div
FEI Number
3000210122
Product Code
KDJ
Status
Terminated
Root Cause
Other
Initiated
November 5, 2004
Posted
December 9, 2004
Terminated
May 16, 2006
Address
1620 Waukegan Rd, Bldg R, Mc Gaw Park, IL, 60085-6730

Description

Baxter CAPD Solution Transfer Set for use with UV-Flash Germicidal Exchange Device, 1.2 m (48''); a sterile set consisting of a bag connector (spike connector), on-off clamp assembly, tubing and double sealing male Luer lock connector; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085; product code R5C4325

Reason

Disconnection of the tubing at the clamp level of the UV Flash Transfer Set.

Action

Recall by letter dated 11/05/04. The customers were instructed to stop using this lot of product, return any stocks of the lot back to Baxter, and that replacement of the patient''s set be completed at the dialysis center.

Distribution

The entire lot was distributed overseas to Belgium, Sweden, Norway, the United Kingdom, Germany and France.

Quantity

1,152 sets