24CM DECATHLON FUNNEL TIP CATHETER Model: DF19SH24-100 DECATHLON FUNNEL TIP - CATHETER ONLY 15.5 Fr polyurethane twin lumen hemodialysis catheter with separated tips, cuff and side holes for chronic hemodialysis and apheresis. Latex free.
Recall
- Recall Number
- Z-0300-2010
- Event Number
- 53605
- Firm
- Spire Biomedical, Inc. One Patriots Park
- FEI Number
- 1000159376
- Product Code
- MSD
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- October 23, 2009
- Posted
- November 17, 2009
- Terminated
- June 9, 2010
- Address
- Bedford, MA, 01730-2343
Description
24CM DECATHLON FUNNEL TIP CATHETER Model: DF19SH24-100 DECATHLON FUNNEL TIP - CATHETER ONLY 15.5 Fr polyurethane twin lumen hemodialysis catheter with separated tips, cuff and side holes for chronic hemodialysis and apheresis. Latex free.
There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira
Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.
Worldwide distribution - Nationwide, Government. Foreign: Canada, France.
36 units