FDA Recall Terminated

Ysio (model number 10281013 or 10281163) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used for radiographic and tomographic exposures of the whole body.

Recall: Z-0299-2019 · Initiated August 6, 2018

Recall

Recall Number
Z-0299-2019
Event Number
80960
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
KPR
Status
Terminated
Root Cause
Device Design
Initiated
August 6, 2018
Terminated
April 10, 2020
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

Ysio (model number 10281013 or 10281163) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used for radiographic and tomographic exposures of the whole body.

Reason

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.

Action

Urgent Medical Device Correction notification letters dated 8/6/18 were distributed to customers. The letters instruct customers to do the following: To avoid the above mentioned risk until implementation of the corrective action, it is strongly recommended not to perform up/down movements of the lifting column directly above the patient. Instead, the vertical movement should be completed beside the patient and then the system should be moved horizontally to the patient. How will the issue finally be resolved? Siemens Healthineers is preparing a corrective action that will resolve this potential issue. The corrective action will be distributed via an Update Instruction XP025/18/S and is planned to be available in the end of August 2018. How will the corrective action be implemented and how effective is the corrective action? Our service organization will contact you shortly to arrange a date to perform this corrective action with Update Instruction XP025/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.

Distribution

The products were distributed to the following US states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, and WY.

Quantity

142