FDA Recall Terminated

40CM DECATHLON FUNNEL TIP STD KIT Model: DF35SH40 DECATHLON FUNNEL TIP 15.5 Fr polyurethane twin lumen hemodialysis catheter with separated funnel tips, cuff and side holes for chronic hemodialysis and apheresis. Kit contains catheter with stylet, 18 gauge introducer needle, 12cc syringe, J/straight 0.038 inch guidewire, 12Fr dilator, 14Fr dilator, 16Fr peel-away sheath/dilator, #11 scalpel, tunneling stylet, (4) gauze sponges, wound dressing, (2) injection sealing caps, and warning label. Latex free.

Recall: Z-0298-2010 · Initiated October 23, 2009

Recall

Recall Number
Z-0298-2010
Event Number
53605
Firm
Spire Biomedical, Inc. One Patriots Park
FEI Number
1000159376
Product Code
MSD
Status
Terminated
Root Cause
Process control
Initiated
October 23, 2009
Posted
November 17, 2009
Terminated
June 9, 2010
Address
Bedford, MA, 01730-2343

Description

40CM DECATHLON FUNNEL TIP STD KIT Model: DF35SH40 DECATHLON FUNNEL TIP 15.5 Fr polyurethane twin lumen hemodialysis catheter with separated funnel tips, cuff and side holes for chronic hemodialysis and apheresis. Kit contains catheter with stylet, 18 gauge introducer needle, 12cc syringe, J/straight 0.038 inch guidewire, 12Fr dilator, 14Fr dilator, 16Fr peel-away sheath/dilator, #11 scalpel, tunneling stylet, (4) gauze sponges, wound dressing, (2) injection sealing caps, and warning label. Latex free.

Reason

There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira

Action

Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.

Distribution

Worldwide distribution - Nationwide, Government. Foreign: Canada, France.

Quantity

43 units