FDA Recall Terminated

Siemens, SONOLINE Antares Ultrasound System, 5.0 Systems with software version 200.0.054, P/N 05936518 and 10037593 , Siemens, Ultrasound Division, Mountain View, CA 94039

Recall: Z-0298-2008 · Initiated July 10, 2007

Recall

Recall Number
Z-0298-2008
Event Number
45511
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
2936884
Product Code
IYN
Status
Terminated
Root Cause
Software design
Initiated
July 10, 2007
Posted
January 9, 2008
Terminated
December 20, 2010
Address
1230 Shorebire Way P.O. Box 7393, Mountain View, CA, 94043

Description

Siemens, SONOLINE Antares Ultrasound System, 5.0 Systems with software version 200.0.054, P/N 05936518 and 10037593 , Siemens, Ultrasound Division, Mountain View, CA 94039

Reason

Image error: A software problem results in on-screen indications (graphics) that lead the user to believe that the patient's right and left and the transducer orientation are oriented the same. This is not the case and the actual patient's right and left are reversed from the on-screen indicators.

Action

The firm sent a recall letter, Urgent Medical Device Notification, to consignees on 07/10/2007, to be followed by a service visit by Siemens personnel for a mandatory software update to be installed. Short term the users were instructed to avoid increasing the SieClear compounding levels to an Advanced SieClear compounding setting after an L/R invert action with Biopsy guidelines on-screen.

Distribution

Worldwide: Product was distributed to medical facilities Nationwide including the states of OH, FL, IL, NY, CA, NH, TX, NJ, PA, MD, MO, GA, WA, WI, NE, WV, HI, MA, SC, MI, VA, AZ, UT, NH, RI, KY, SD, and IN, and internationally to Canada, Germany, Australia, Sweden, Saudi Arabia, Chile, South Korea, Mexico, France, UK, Taiwan, Singapore, Denmark, Brazil, New Zealand, Ireland, Belgium, Norway, Serbia, Vietnam, UK, Poland, China, Spain, South Africa, Austria, Croatia, Russia, Greece, Trinidad and Tobago, United Arab Emirates, India, Japan, Malaysia, Turkey, Thailand, Norway, Slovakia, Romania, Netherlands, Switzerland, Portugal, Italy, and Venezuela.

Quantity

418 units (Z-0297-0298-2008)