FDA Recall
Terminated
Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.
Recall: Z-0298-04
·
Initiated November 26, 2003
Recall
- Recall Number
- Z-0298-04
- Event Number
- 27905
- Firm
- Ortho-Clinical Diagnostics
- FEI Number
- 1000136573
- Product Code
- JHX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 26, 2003
- Posted
- January 6, 2004
- Terminated
- February 20, 2004
- Address
- 100 Indigo Creek Drive, Rochester, NY, 14650-0881
Description
Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.
Reason
Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
Action
Customers notified via phone and faxed letter on 11/26/03. Customers provided with a modified QC procedure to be used until new shipments can be made.
Distribution
738 reagent packs shipped to U.S. consignees and 36 reagent packs to Canadian consignees.
Quantity
774 reagent packs