FDA Recall Terminated

Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.

Recall: Z-0298-04 · Initiated November 26, 2003

Recall

Recall Number
Z-0298-04
Event Number
27905
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
JHX
Status
Terminated
Root Cause
Other
Initiated
November 26, 2003
Posted
January 6, 2004
Terminated
February 20, 2004
Address
100 Indigo Creek Drive, Rochester, NY, 14650-0881

Description

Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.

Reason

Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.

Action

Customers notified via phone and faxed letter on 11/26/03. Customers provided with a modified QC procedure to be used until new shipments can be made.

Distribution

738 reagent packs shipped to U.S. consignees and 36 reagent packs to Canadian consignees.

Quantity

774 reagent packs