FDA Recall Terminated

AxSYM Digoxin II Reagent Pack, Disk Version 1. (3D53-01 or 3D53-02) The AxSYM Digoxin II assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of digoxin, a cardiovascular drug, in serum or plasma. The measurements obtained are used in the treatment of digoxin overdose and monitoring levels of digoxin to ensure appropriate therapy.

Recall: Z-0297-2009 · Initiated May 22, 2008

Recall

Recall Number
Z-0297-2009
Event Number
48628
Firm
Abbott Diagnostic International, Ltd. Carr # 2 Km 58.0 Cruce Davila
FEI Number
2623532
Product Code
KXT
Status
Terminated
Root Cause
Software change control
Initiated
May 22, 2008
Posted
November 21, 2008
Terminated
September 9, 2011
Address
Barceloneta, PR, 00617-3009

Description

AxSYM Digoxin II Reagent Pack, Disk Version 1. (3D53-01 or 3D53-02) The AxSYM Digoxin II assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of digoxin, a cardiovascular drug, in serum or plasma. The measurements obtained are used in the treatment of digoxin overdose and monitoring levels of digoxin to ensure appropriate therapy.

Reason

AxSYM TDM/ Transplant Assay Disks 3D53-03 and higher do not contain the AxSYM Digoxin II Assay file. The instrument might require installation of the AxSYM Digoxin II assay file.

Action

A Product Information letter, dated May 22, 2008, was issued to customers. It requested that one of the following two actions be taken if the AxSYM Digoxin II assay file is not installed on your AxSYM instrument and you want to run the AxSYM Digoxin II assay. If you have one of the previously manufactured software disks listed, [AxSYM Digoxin II Assay Disk, LN 2C43-01, or higher; or AxSYM TDM/Transplant Assay Disk Version 1.0 (3D53-01) or Version 2.0 (3D53-02)], install the AxSYM Digoxin II assay file. Do not install any other assay file. If you do not have one of the assay disks listed, contact your local Customer Service representative to have this assay file installed or install the file from a previous version if you have it on hand.

Distribution

Worldwide Distribution --- USA including states of AL, AZ, AR, CA, CT, FL, GA, IL, KS, MD, MA, MI, MN, MS, MO, NV, NJ, NY, ND, OK, OR, PA, RI, TX, VA, WA, Hawaii, and Puerto Rico, and countries of Canada, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Ecuador, Peru, Chile, Uruguay, Argentina, Puerto Rico, Germany, Japan, Singapore, Hong Kong, Taiwan, Thailand, South Korea, Australia, New Zealand, Curacao, and Panama.

Quantity

9069 kits