St. Jude Medical Merlin PCS Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3650
Recall
- Recall Number
- Z-0297-2007
- Event Number
- 36619
- Firm
- St Jude Medical CRMD
- FEI Number
- 2017865
- Product Code
- KRG
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 6, 2006
- Posted
- December 19, 2006
- Terminated
- March 26, 2012
- Address
- 15900 Valley View Ct, Sylmar, CA, 91342-3577
Description
St. Jude Medical Merlin PCS Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3650
St. Jude Medical has identified a low-frequency anomaly in the software used in the APS III Model 3500/3510 and Merlin PCS Model 3650 programmers that can lead to incorrect reporting of battery voltage, expected battery longevity and Elective Replacement Indicator (ERI) status.
The notification material was sent to physicians by FedEx. The advisory notification was sent by FedEx on Oct. 12, 2006, to arrive at the physicians' offices on Oct. 13th. In the US, acknowledgement of receipt of the notification material will be monitored by St. Jude Medical CRMD in Sylmar, California. For the rest of the world it will be by various St. Jude offices and local distributors with this information being fed back to St. Jude Medical CRMD in Sylmar. In addition, St. Jude Medical issued a press release on October 13, 2006. Plus information concerning this advisory can be found on St. Jude's website, www.sim.com Completion of this advisory is estimated to be three months after its initiation.
Worldwide in USA, ARGENTINA, AUSTRALIA, AUSTRIA, BARBADOS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, ESTONIA, EUROPE, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GUATEMALA, HONG KONG, ICELAND, INDIA, ISRAEL, ITALY, JAPAN, JORDAN, SOUTH KOREA, LATVIA, LICHTENSTEIN, LUXEMBURG, MALAYSIA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, PORTUGAL, SAUDI ARABIA, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, UNITED KINGDOM & VENEZUELA
6,290