MagNA Pure LC Instrument; catalog numbers 2236931 and 03670325001.
Recall: Z-0297-05
·
Initiated November 5, 2004
Recall
- Recall Number
- Z-0297-05
- Event Number
- 30483
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- NPQ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 5, 2004
- Posted
- December 8, 2004
- Terminated
- August 29, 2005
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
MagNA Pure LC Instrument; catalog numbers 2236931 and 03670325001.
Reason
Test results may be adversely affected by software shutting down the cooling blocks in systems with software version 3.09.
Action
Consignees were notified via letter dated 11/5/04, which provides the user with a workaround.
Distribution
Nationwide.
Quantity
65