Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratories, Abbott Park, IL 60064
The Operations Manual has the wrong Farenheit temperature range listed (102-122) instead of (104-122) for the preparation of the Terg-A-Zyme solution used to clean pipettes of the Abbott Commander Flexible Pipetting Center.
Abbott sent 'Important Product Information' letters dated 12/2/03 to the end user accounts, advising them of the error in conversion of the Celsius temperature range to Farenheit in the first step in preparing the Terg-A-Zyme cleaning solution. The letter listed the entire corrected solution preparation procedure and advised the accounts to retain the letter with the Abbott FPC Operations Manual for future reference.
Nationwide and internationally through Abbott affiliates in Colombia, Brazil, Puerto Rico, Japan, Singapore, Hong Kong and Germany.