Terumo Sarns Level Sensor II (red) (Part # 195274) alarm level sensor transducer for use with Sarns Advanced Perfusion System I, Perfusion System 800 and Perfusion System 900; Terumo Cardiovascular Systems Corp., Ann Arbor, MI.; REF 195274. Level sensors are distributed as a finished device, as part of a kit, with a perfusion system 8000 safety monitor or with an Advanced Perfusion System 1 base. Level sensors are attached to the color coded connectors on the module and then to a hard-shell reservoir. Level detection is used to monitor blood levels in hard-shell reservoirs.
Recall
- Recall Number
- Z-0296-2009
- Event Number
- 49604
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DTQ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 23, 2008
- Posted
- November 7, 2008
- Terminated
- August 4, 2009
- Address
- 6200 Jackson Rd, Ann Arbor, MI, 48103
Description
Terumo Sarns Level Sensor II (red) (Part # 195274) alarm level sensor transducer for use with Sarns Advanced Perfusion System I, Perfusion System 800 and Perfusion System 900; Terumo Cardiovascular Systems Corp., Ann Arbor, MI.; REF 195274. Level sensors are distributed as a finished device, as part of a kit, with a perfusion system 8000 safety monitor or with an Advanced Perfusion System 1 base. Level sensors are attached to the color coded connectors on the module and then to a hard-shell reservoir. Level detection is used to monitor blood levels in hard-shell reservoirs.
The level sensor may not properly couple to the reservoir, resulting in a sensor "not attached" message, an alert or alarm condition, or a failure to detect a low level condition.
U.S. consignees were notified by (Urgent Medical Device Removal) letter dated 5/23/08 via FedEx and international affiliates were notified via e-mail on 5/27/08. Consignees were advised of the problem, and to continue using the affected sensors until replacements are available in late 2008. For questions or additional information, contact the Terumo CVS Customer Service at 1-800-521-2818.
Worldwide including USA and countries of Australia, Belgium, Bolivia, Canada, Chile, Ecuador, Egypt, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Latin America, Malaysia, Panama, Philippines, Saudi Arabia, Singapore, South Africa, South Korea, Sri Lanka, Taiwan, Thailand, United Arab Emirates, Venezuela and Vietnam.
547